Is the Elapsed Time following the Placement of a Ventriculoperitoneal Shunt Catheter an Individual Risk Factor for Shunt Fractures?

[ X ]

Tarih

2012

Dergi Başlığı

Dergi ISSN

Cilt Başlığı

Yayıncı

Karger

Erişim Hakkı

info:eu-repo/semantics/closedAccess

Özet

In this study, we examined whether the resistance of peritoneal catheters against the retraction force changed over time following shunt placement, and the role of this resistance in shunt fracture is discussed. We investigated peritoneal catheters removed from patients treated with a ventriculoperitoneal shunt because of hydrocephalus; previously, patients underwent shunt revision. The maximum tension, maximum elongation and elongation percentages of the peritoneal catheters were measured. The mean and maximum tension values of the revised peritoneal catheters were increased compared to the unused catheters. The maximum elongation and elongation rates were significantly decreased. The changes in the maximum elongation, elongation rate and tension values were unrelated to the time elapsed after catheter insertion. This finding indicates that the time elapsed following peritoneal catheter placement was not an individual factor based on the strength of the response of the organism to the foreign body and the mechanical trauma exposed in shunt fractures. Copyright (C) 2013 S. Karger AG, Basel

Açıklama

Anahtar Kelimeler

Catheter, Cerebrospinal Fluid Shunt, Hydrocephalus, Shunt Complications

Kaynak

Pediatric Neurosurgery

WoS Q Değeri

Q4

Scopus Q Değeri

Q3

Cilt

48

Sayı

6

Künye