Comparison of the accuracy of different electronic apex locators used in working length determination via changing of initial cleaning solutions

dc.authorid0000-0003-3116-2929en_US
dc.authorid0000-0002-4644-0058en_US
dc.contributor.authorAvcı, Güliz Rana Tellioğlu
dc.contributor.authorKaya, Sadullah
dc.date.accessioned2022-10-03T12:06:47Z
dc.date.available2022-10-03T12:06:47Z
dc.date.issued2021en_US
dc.departmentDicle Üniversitesi, Diş Hekimliği Fakültesi, Endodonti Bölümüen_US
dc.description.abstractAim: The aim of this study is to compare the accuracy of the different new generation electronic apex locators by changing the initial cleaning solutions (NaOCl vs. EDTA). Methodology: Eighty mandibular premolar teeth were used. Radiographs were obtained from the mesio-distal, and bucco-lingual angles for all teeth, and teeth with suspected external or internal root resorption and/or calcified canals were not included in the study. The true lengths of the root canals were determined with a precision of 0.01 mm with a stereomicroscope at x5 magnification by placing a #20 K-type canal file into the root canal. The samples were then embedded in freshly mixed alginate blocks up to the cemento-enamel junction. The coronal 1/3 was pre-enlarged with a Protaper SX (VDW, Munich, Germany), instrument. Root ZX Mini (J. Morita, Tokyo, Japan), Raypex 6 (VDW, Munich, Germany), Propex Pixi (Dentsply Maillefer, Ballaigues, Switzerland), and E-Pex Pro (Changzhou Sifary Medical Technology Co., Ltd, Jiangsu, China) electronic apex locators were used, and measurements were made by changing the initial solutions (NaOCl vs. EDTA). Distilled water was used between application of 5.25% NaOCl and 17% EDTA to prevent chemical interactions. Three measurements were taken for each tooth, and the average of these three measurements was taken as reference. The data obtained in this study were analyzed. Results: There were no statistically significant differences between the solutions in terms of the measurement values and actual length values found by each device (p>0.05). Although there were no statistically significant differences, the deviation from the actual measurement was greater when EDTA solution was used than when NaOCl solution was used. In addition, although the difference was not found to be statistically significant, the deviation from the actual measurement was found to be highest when the Root ZX mini device was used with NaOCl and EDTA solution. Conclusion: The results of this study showed that the difference in the initial washing solutions did not affect the electronic apex locator (EAL) devices in making measurements close to the true length.en_US
dc.identifier.citationAvcı, G. R. T. ve Kaya, S. (2021). Comparison of the accuracy of different electronic apex locators used in working length determination via changing of initial cleaning solutions. International Dental Research, 11(Supplement 1), 56-62.en_US
dc.identifier.doi10.5577/intdentres.2021.vol11.suppl1.9
dc.identifier.endpage62en_US
dc.identifier.issn2146-1767
dc.identifier.issueSupplement 1en_US
dc.identifier.startpage56en_US
dc.identifier.trdizinid516630
dc.identifier.urihttps://search.trdizin.gov.tr/yayin/detay/516630
dc.identifier.urihttps://hdl.handle.net/11468/10599
dc.identifier.urihttps://search.trdizin.gov.tr/yayin/detay/516630
dc.identifier.volume11en_US
dc.indekslendigikaynakTR-Dizin
dc.institutionauthorAvcı, Güliz Rana Tellioğlu
dc.institutionauthorKaya, Sadullah
dc.language.isoenen_US
dc.publisherUluslararası Diş Araştırmaları Birliğien_US
dc.relation.ispartofInternational Dental Research
dc.relation.publicationcategoryMakale - Uluslararası Hakemli Dergi - Kurum Öğretim Elemanıen_US
dc.rightsinfo:eu-repo/semantics/openAccessen_US
dc.subjectWorking lengthen_US
dc.subjectElectronic apex locatoren_US
dc.subjectInitial solutionsen_US
dc.titleComparison of the accuracy of different electronic apex locators used in working length determination via changing of initial cleaning solutionsen_US
dc.titleComparison of the accuracy of different electronic apex locators used in working length determination via changing of initial cleaning solutions
dc.typeArticleen_US

Dosyalar

Orijinal paket
Listeleniyor 1 - 1 / 1
Yükleniyor...
Küçük Resim
İsim:
Comparison of the accuracy of different electronic apex locators used in working length determination via changing of initial cleaning solutions.pdf
Boyut:
428.38 KB
Biçim:
Adobe Portable Document Format
Açıklama:
Makale Dosyası
Lisans paketi
Listeleniyor 1 - 1 / 1
[ X ]
İsim:
license.txt
Boyut:
1.44 KB
Biçim:
Item-specific license agreed upon to submission
Açıklama: