THERMAL RESPONSE TO MULTIPLE USE OF A TWIST DRILL

[ X ]

Tarih

1992

Dergi Başlığı

Dergi ISSN

Cilt Başlığı

Yayıncı

Mosby-Year Book Inc

Erişim Hakkı

info:eu-repo/semantics/closedAccess

Özet

The objective of the present investigation was to determine whether injurious thermal changes occur in the pulp chamber if a twist drill is used 10 times, and to find whether there is a correlation between temperature rise and distance between a drilled channel and the pulp chamber. Twenty caries-free, human, premolar teeth were randomly distributed to two dentists. With a new 2 mm twist drill, each dentist prepared 10 pinholes (one hole per tooth). Intrapulpal temperature change during the drilling procedure was recorded. The data were statistically analyzed by the Spearman rank order correlation coefficient. The results showed that twist drills should not be used for more than five pinholes (p < 0.001). The distance between the pin channel and the pulp chamber does not influence heat generation in the pulp chamber.

Açıklama

Anahtar Kelimeler

[No Keyword]

Kaynak

Journal of Prosthetic Dentistry

WoS Q Değeri

N/A

Scopus Q Değeri

Q1

Cilt

67

Sayı

4

Künye