Optical and mechanical clot detection methodologies: A comparison study for routine coagulation testing

Yükleniyor...
Küçük Resim

Tarih

2012

Dergi Başlığı

Dergi ISSN

Cilt Başlığı

Yayıncı

Erişim Hakkı

info:eu-repo/semantics/openAccess

Özet

Background: Automated coagulation analyzers are preferred to meet increasing coagulation test volume. Two distinct technological families exist based on optical and mechanical clot detection methodologies. Which one is superior to the other is still a conflict and needs new studies. Methods: We have compared prothrombin and activated partial thromboplastin results obtained with mechanical method with those obtained by photo-optical method used routinely in our specialized laboratory. Results: The instrumental results showed good precision ranging between 0.7% and 1.8% coefficient of variation. Statistical analysis demonstrated an excellent correlation between the photo-optical and mechanical analyzers for PT (R2 0.97), and aPTT (R2 0.85). Conclusion: Correlation between the two clot-detection systems was maintained even when measuring turbid samples (R2 ≥ 0.97 for two tests).

Açıklama

Anahtar Kelimeler

Blood coagulation, Clot detection, Factors, Mechanical detection, Optical detection, Partial thromboplastin time, Prothrombin time

Kaynak

Journal of Clinical Laboratory Analysis

WoS Q Değeri

Q3

Scopus Q Değeri

Q1

Cilt

26

Sayı

3

Künye

Tekkesin, N. ve Kılınç, C. (2012). Optical and mechanical clot detection methodologies: A comparison study for routine coagulation testing. Journal of Clinical Laboratory Analysis, 26(3), 125-129.