Cardiac device-related infective endocarditis; analysis of 15 cases

[ X ]

Tarih

2013

Dergi Başlığı

Dergi ISSN

Cilt Başlığı

Yayıncı

Turkish Soc Cardiology

Erişim Hakkı

info:eu-repo/semantics/openAccess

Özet

Objectives: We aimed to investigate the demographic and clinical characteristics, echocardiographic and microbiologic features, and outcomes of patients with permanent pacemaker (PM), and implantable cardioverter-defibrillator (ICD) - related endocarditis in this study. Study design: The study population consisted of 15 patients with permanent PM and ICD-related endocarditis. Data on patients'demographic characteristics, medications used, clinical, and microbiological data, echocardiographic findings, types, and outcomes of surgical treatments were recorded. Results: The mean age of the patients was 57+/-16. Seven patients (47%) were female. Of the 15 patients with permanent PM and ICD-related endocarditis, 5 died during in-hospital follow-up (33%). In four patients (27%) pulmonary embolism developed. Culture-negative endocarditis was detected in 5 cases (33%). Staphylococci were the most common causative organisms in 60% of the patients. Three (20%). patients underwent surgical treatment Conclusion: Cardiac device-related endocarditis remains a rare complication of intracardiac device implantation still with higher mortality rates.

Açıklama

Anahtar Kelimeler

Defibrillators, Implantable/Adverse Effects, Endocarditis/Diagnosis/Prevention & Control /Therapy, Pacemaker, Artificial, Cardiology

Kaynak

Turk Kardiyoloji Dernegi Arsivi-Archives of The Turkish Society of Cardiology

WoS Q Değeri

N/A

Scopus Q Değeri

Q3

Cilt

41

Sayı

2

Künye