Comparison of Central Corneal Thickness Measurements with a Rotating Scheimpflug Camera, a Specular Microscope, Optical Low-Coherence Reflectometry, and Ultrasound Pachymetry in Keratoconic Eyes

[ X ]

Tarih

2015

Dergi Başlığı

Dergi ISSN

Cilt Başlığı

Yayıncı

Taylor & Francis Inc

Erişim Hakkı

info:eu-repo/semantics/closedAccess

Özet

Objective: To compare central corneal thickness (CCT) measurements with a rotating Scheimpflug camera (RSC), noncontact specular microscopy (SM), optical low-coherence reflectometry (OLCR), and ultrasonic pachymetry (UP) in keratoconus (KC) patients. Method: In this prospective study, four CCT measurements taken with an RSC, SM, OLCR, and UP were compared in 81 eyes of 44 consecutive KC patients. The KC patients were divided into four subgroups according to Amsler-Krumeich's KC classification. Results: The RSC and UP measurements of the CCT were not statistically significant in all the groups. Comparison of the SM vs. the OLCR measurements yielded statistically significant differences in all the KC patients and in all KC stages. In all the KC patients, RSC and OLCR showed a high correlation coefficient factor (r = 0.87, p = 0.000). Conclusion: CCT measurements with RSC are comparable to those achieved with UP. Compared with the other devices, according to SM measurements, the central cornea is thicker in all keratoconic eyes and in all KC grades, and it is thinner according to OLCR. RSC, UP, SM, and OLCR should not be used interchangeably in keratoconic eyes.

Açıklama

Anahtar Kelimeler

Central Corneal Thickness, Keratoconus, Pachymetry, Scheimpflug Camera, Reflectometry

Kaynak

Seminars in Ophthalmology

WoS Q Değeri

Q4

Scopus Q Değeri

Q2

Cilt

30

Sayı

2

Künye